VisionR/S Series - Common Troubleshooting

VisionR/S Series - Common Troubleshooting

During refraction or usage of the device, the keyboard is very slow to react 

  1. Check the % on the top left corner

  2. If it shows less than 40%  - Switch the Vision-R system OFF and back ON.

  3. Remove fingerprint smudges with a tissue or soft microfiber cloth.

  4. Check that the B letter is displayed temporarily in the top left corner when doing any action

  5. If the error message is still displayed even after performing the procedure above, turn off the power, disconnect the power cord and contact technical support.



The objective data can not be imported or exported. 

  1. Error code: 1005 – Failed to communicate with CBOX

  2. Ensure that the internet connection is on and working in the office.  

  3. If the error message is still displayed even after performing the procedure above, turn off the power, disconnect the power cord and contact 


The Monitor has gone black.

  1. The sleep function may be activated. 

  2. Press any button to wake the device from the sleep mode.



Nothing appears on the keyboard display when the power switch is turned on. Double-check all power connections for the keyboard, phoropter head, and power supply module first before proceeding.

  1. Case n°1: Blue-touch button on the top right-hand side of the keyboard is illuminated 

    1. Turn the keyboard OFF and back ON.  Watch and listen to the phoropter head to see if the initialization process occurs.   

    2. •If the initialization does not occur and no sound is heard from the phoropter head, there may be an issue with the power supply. Contact technical service.

    3. •If the initialization process of the phoropter head occurs.. Make sure the RJ45 cable is secured to the Internet port of the power supply.

  2. Case n°2: Blue-touch button on the top right-hand side of the keyboard is not illuminated

    1. It is likely that a power supply cord is not secure. Check the power cord from the keyboard to the power supply and the power supply to the wall to ensure that they are secure and well seated. 

  3. Case n°3 : Blue-touch button on the top right-hand side of the keyboard lights up then turns OFF

    1. Unplug the cable of the head and start the process again

    2. If the issue still persists, contact customer service



When selecting an optotype, it is not displayed on CSPola600 

  1. The Vision-R 800 communicates with the CSPola600 screen via Infrared. There are several things that can cause this: 

  2. Ensure that the CSPola600 screen is turned on and plugged in securely to the power outlet. 

  3. A halogen bulb light source in the room can affect the Infrared signal between the phoropter and the chart screen

    1. Turn the light source OFF and change the test. Then turn the light source back ON to proceed.

    2. If the issue is resolved when the halogen light is turned off, it is recommended to change to LED bulbs. 

  4. The room has large frosted surfaces made up of glass panels that reflect the InfraRed signal. 

    1. Change the position/orientation of the CSPola &/or of the head 

  5. Ensure that there is no smartphone or tablet with a TrueDepth camera (3D) near the devices.. The technology used for this affects the Infrared signal.. 



Impossible to adjust PD, Vertex or change data on one eye (ERROR 1023)

  1. Error due to eye lock (left, right or both) is displayed at anytime you want to modify

  2. Unlock the keyboard. 


      



One of the following messages are displayed when performing the exam :


“The total sphere is out of range. “

The Total cylinder value is out of range.”

“The sphere-cylinder combination is out of range. “

“The prism value is out of range. “



  1. This message appears in case of the value exceeds the upper limit of the

measurement range of the device.

  1. Take a measurement within the measurement range. 

  2. It will also occur if your system is locked from doing -/+ Cylinder,  -/+ sphere, or -Add.

  3. Sphere :-20D to +20D

Cylinder : -8D to +8D

Prism-20D to +20D.











    Submit a ticket


      Remote Support







        • Related Articles

        • VisionR/S Series - Keys

          Locate each button.  1. Clear Data   Short press = Current values reset. Long press = Switch the instrument on/off 2. Import   Import patient’s refraction data. 3. Export   Export patient’s refraction data 4. Far/Near Vision :  Far vision mode ...
        • VisionR/S Series - Near Vision

          1 Jackson Cross : Smart Test & Standard test Determine the addition 2 Acuity letters  C Landolt Evaluating Near Acuity 3 Dots chart  Dial clock Checking Astigmatism 4  Row of vertical letters Lines of horizontal letters Evaluating phorias (Van ...
        • VisionR Series - Adjusting the level

          Follow this recommendation to manipulate the VisionR    STEP 1    Introduce the exam to the patient  “This device is the Vision-R® 800. This breakthrough device will enable us to measure your refraction at a 0.01 diopter, the highest level of ...
        • VisionR/S Series - Measuring the Vertex

          STEP 1 Enter the Vertex distance mode " Please look straight ahead and open your eyes. " STEP 2 Adjust the blue line(s) until it is marking the cornea apex on the right and left eyes, using the dial jog.  (Monocular or Binocular adjustment).  STEP 3 ...
        • VisionR/S Series - Touch Screen

          Locate each section and working areas .              
        • Recent Articles

        • How to remove the IPL cartridges?

          To remove any of the terminals, hold the hand piece with both hands. Place the thumb under the terminal and push up to disengage. Use your other hand to slide the terminal out.
        • How to remove the IPL cartridges?

          To remove any of the terminals, hold the hand piece with both hands. Place the thumb under the terminal and push up to disengage. Use your other hand to slide the terminal out.
        • How to remove the IPL cartridges?

          To remove any of the terminals, hold the hand piece with both hands. Place the thumb under the terminal and push up to disengage. Use your other hand to slide the terminal out.
        • How to remove the IPL cartridges?

          To remove any of the terminals, hold the hand piece with both hands. Place the thumb under the terminal and push up to disengage. Use your other hand to slide the terminal out.
        • What are the benefits of LM LLT?

          It is fast—a treatment lasts 15 minutes. It is painless. It grants immediate relief to the patient. It is easy and safe for the operator. It is plug-and-play and does not require the operator to be constantly present during the treatment.